Direct Measurement of Dust Concentration by Shadowgraphy Method—Direct Image Particle Analysis (DIPA)

SHENG-CHIEH CHEN, Seungkoo Kang, Wing-Tak Lai, David Y. H. Pui, Virginia Commonwealth University

     Abstract Number: 1670
     Working Group: Instrumentation

Abstract
ISO dust injectors are generally applied to re-suspend standard dust particles, e.g. ISO A2, to challenge air filters used in internal combustion engines. The size distribution of the re-suspended dust particles is very important for acquiring the size fractional efficiency and loading performance of the filter. Generally, the existing size distributions provided by the powder manufacturer are directly applied or real-time aerosol instruments are used to
measure the particles. Various instruments, however, report different size distributions for the same dust sample and also they are quite different with that provided by the manufacturer. The main reasons are likely due to the particle transport loss during the sampling by the real-time instruments, especially for dust particles larger than 1 μm, as well as the different measurement principles and different sizing ranges of the instruments. In addition, the provided size distributions by the manufacturer are not ideal to represent the re-suspended dust particles because those distributions are measured in liquid when they are highly deagglomerated by the aids of chemical agents and ultra-sonication. In this study, therefore, the in-situ and noninvasive shadowgraph technique with an image analysis technique (particle/droplet image analysis, PDIA) were applied to measure the size of the re-suspended dusts at the moment they are being generated. The experimental system consisted of an 8 Mpixel CCD camera equipped with a high magnification micro-lens, up to 28×, to allow the measurement of small particles down to 1 μm. Monodisperse PSL particles with diameters of 5, 17 to 26 μm were generated from a homebuilt generator and used to validate the sizing accuracy of the system. Besides, the standard sized dots with 1, 3.5, 10, 50 and 100 um were used to calibrate the depth of field of the system. The validated system was then applied to measure size distribution of the widely used ISO A2 fine dusts re-suspended by different dust dispersers, including the ISO light-duty and ISO heavy-duty injectors. Results showed that there was a large discrepancy between the size distributions determined by the powder manufacturer and those from ISO injectors by PDIA, indicating the ISO injectors did not deagglomerate A2 dustswell. This study first reported the reliable size distributions for A2 dusts re-suspended by ISO injectors under different operation conditions. In addition to A2 dusts, the known sizes and known concentrations PSL particles suspended in DI water were also characterized. It was found the PDIA was able to not only accurately measure particle sizes but also directly measure particle concentration. This method should be widely applied for measuring in-situ size distribution of dust particles used to challenge the filters as well as particles in liquid.